HANNA HI9126 Calibration Check Portable pH/ ORP Meter

HANNA HI9126 Calibration Check Portable pH/ ORP Meter

Rp. 21,500,000
NOTE
Harga dapat berubah sewaktu-waktu mengikuti perkembangan kurs rupiah terhadap mata uang asing lainnya sehingga sangat mungkin terjadi adanya perbedaan harga yang berlaku dengan harga yang tertera pada website kami.

The HI 9126 includes HANNA‘ s exclusive Calibration Check™ technology. Calibration Check™ monitors the pH bulb and reference junction of the electrode every time the instrument is calibrated. In the event of a dirty pH electrode, Calibration Check™ warns users that maintenance may be needed.

Calibrated buffers are continuously displayed in measurement mode to remind users of the instrument’ s calibration point. Users can easily determine if readings are taken too far outside the calibration range.

The HI 9126 can store and recall readings at the touch of a button.
HI 9126 utilizes the HI 1230B double junction pH electrode. The double junction helps to minimize junction contamination for consistently accurate results. The HI 9126 can also measure ORP using optional probes.

Features: 
pH Calibration Check™ 
Electrode condition monitoring
Measurement store and recall
Backlit, multi-level LCD display
Real time clock
User-selectable calibration reminder
Tutorial messages displayed on LCD
BEPS; ( Battery Error Prevention System) alerts the user in the event that low battery power could adversely affect readings.
Battery % displayed on startup.

Specification
Range    pH    -2.00 to 16.00 pH
ORP    ±699.9 mV; ±1999 mV
Temperature    -20.0 to 120.0°C / -4.0°F to 248.0°F
Resolution    pH    0.01 pH
ORP    0.1 mV; 1 mV
Temperature    0.1°C / 0.1 °F
Accuracy @ 20°C    pH    ±0.01 pH
ORP    ±0.2 mV; ±1 mV
Temperature    ±0.4°C/±0.8°F (excluding probe error)
pH Calibration
Check         yes, check of the electrode status during calibration
Calibration    pH    automatic, one or two point with seven memorized buffer values (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45) + 2 custom buffers
Offset    ±1 pH
Slope    from 80 to 108%
Temperature
Compensation         automatic or manual, -20 to 120°C (-4 to 248°F) without temperature probe
pH Electrode         HI 1230B PEI body pH electrode with BNC connector and 1 m (3.3’) cable (included)
Temperature Probe         HI 7662 stainless steel temperature probe with 1 m (3.3’) cable (included)
Input Impedance         1012 Ohms
Battery Type / Life         1.5V (3) AAA / approximately 200 hours of continuous use without backlight (50 hours with backlight). auto-off after 20 minutes of non-use (can be disabled)
Environment         0 to 50°C (32 to 122°F); RH max 100%
Dimensions         185 x 72 x 36 mm (7.3 x 2.8 x 1.4”)
Weight         300 g (10.6 oz.)