- Air Quality Meter
- Anemometer
- Battery Test & Analyzer
- Calibration Tester
- Chart & Data Logger
- Chlorine Meter
- Clamp Meter
- Coating Thickness
- Concrete Meter
- Conductivity Meter
- Current & Echosounder
- Current Injection Test
- Current Load Bank
- Deadweight Tester
- Density Meter
- Dew Point Hygrometer
- Dissolved Oxygen
- Fault Finder Locator
- Flaw Detector
- Flow Meter
- Force Gauge
- Gas Analyzer
- Gauss Meter
- Grain Moisture
- Ground Resistance
- Hand Pump
- Hardness Tester
- High Voltage Rercue
- Hipot Tester
- Holiday Detector
- Insulation Tester
- Leak Detection
- Level Lube Controller
- Light Meter
- Manometer
- Multimeter
- Networks Analyzer
- Ohm Meter
- Oil Test Analyzer
- Oscilloscope
- Other Products
- PH Meter
- Pipe Cable Locator
- Power and Voltage
- Power Quality Meter
- Pressure Gauge
- Rangefinder
- Shaft Alignment
- Sound Level Meter
- Spectrophotometer
- Spectrum Analyzer
- Stroboscope
- Tachometer
- Temp. Calibrator
- Thermal Imager
- Thermohygrometer
- Thermometer
- Torque Tester
- Turbidity Meter
- Ultrasonic Thickness
- Vibration Analyzer
- Video Borescope
- Water Level Meter
- Water Quality Meter
- Water Sampler
- Weather Station
- Wood Moisture Meter
OMEGA CL900 Hot Point Dry Block Probe Calibrator
Harga dapat berubah sewaktu-waktu mengikuti perkembangan kurs rupiah terhadap mata uang asing lainnya sehingga sangat mungkin terjadi adanya perbedaan harga yang berlaku dengan harga yang tertera pada website kami.
OMEGA’s new hot point® calibrators allow calibration of thermocouple and RTD probes quickly and easily with accuracy to 0.9°C (1.5°F). An integral RTD sensor assures high measurement accuracy and repeatability. A built-in microprocessor-based PID digital controller assures exceptional stability. Setpoint and actual temperature are displayed simultaneously. Temperature can be set in one-degree increments from 22°C (40°F) above ambient to 482°C (900°F).
The CL900 calibrator accepts a variety of standard and metric optional inserts which are available in 4" and 6" depths and for probe diameters from 1/8" to 3/8". An undrilled insert is also available. The CL900 comes complete with a 1/4" diameter, 6" deep insert well and insert block tongs. Additional inserts can be ordered separately as required for other diameter probes. 6" depth inserts are used for probes 7" and longer; 4" depth inserts are used for probes 5" and longer.
The CL950 calibrator features a built-in well with 5 ports sized to accommodate a 1/16", 1/8", 3/16", and two 1/4" diameter probes. The multi-port design allows for the use of a calibrated or NIST-traceable probe and meter to be used as a standard while calibrating other probes. The 5 port design also allows for calibration of more than one probe at a time, and eliminates the need for separate insert wells.
Specifications
(All specifications are based on the test probes being in contact with the bottom of the test well)
Temperature Range: Ambient +22 to 482°C (+40 to 900°F)
Operating Ambient Temperature Range: 5°C to 38°C (40°F to 100°F)
Accuracy: (1°F resolution)
CL900 with 6" test well inserts: ±1.5°F ±1 LSD.
CL900 with 4" test well inserts: ±3°F ±1 LSD.
CL950: ±1.5°F, ±1 LSD.
Control Stability: ±0.3°F or better
Temperature Uniformity (within the bottom 1" of the test well):
CL900:
with 6" test well inserts: 0.3%
with 4" test well inserts: ±0.4%.
CL950: ±0.3%
Max. Probe Immersion:
CL900:
with 6" test well inserts: 150 mm (6")
with 4" test well inserts: 100 mm (4").
CL950: 150 mm (6")
Control: Microprocessor-based PID digital controller with data retention
Power: 110-125 Vac, 50/60 Hz, 1050 watts; 220 Vac, 50/60 Hz, 1050 watts
Dimensions: 300 H x 216 W x 311 mm D (11.75 x 8.5 x 12.25")
Weight: 10.1 kg (22.3 lb)